Recommended separation distance between portable and mobile RF communications equipment and the
TG120
The TG120 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the TG120 can help prevent electromagnetic interference by main-
taining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the TG120 as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d =1,2√P
80 MHz to 800 MHz
d =1,2√P
800 MHz to 2,5 GHz
d =2,3√P
0,01 N/A 0,12 0,23
0,1 N/A 0,38 0,73
1 N/A 1.2 2.3
10 N/A 3.8 7.3
100 N/A 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
The maximum output power rating of the transmitter in watts (W) according to the transmitter manufac-
turer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorp-
tion and reection from structures, objects and people.
Guidance and manufacturer’s declaration-electromagnetic immunity
The TG120 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG120 should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
Not applicable
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part of
the TG120 including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
ÖZELLİKLER
EMC rehberi ve imalatçının beyanı
GARANTİ
IN0TG12000000000XXXXXXX
OBM_IB_TG120Q_TR_ver1303
Aralık 32.0°C~42.9°C (89.6°F~109.2°F)
Ekran Sıvı kristal ekran, 3 1/2 basamak
Ekran çözünürlüğü 0.1ºC or 0.2ºF
Hassasiyet ±0.1°C 35.5°C~42.0°C (±0.2°F 95.9°F~107.6°F)
±0.2°C altında veya 35.5°C üzerinde 42.0°C
(±0.4°F altında veya 95.9°F üzerinde 107.6°F)
25.0°C (77.0°F)standart oda sıcaklığında .
Bellek Son ölçülen değeri saklamak için
Pil Bir adet 1.5 V DC. düğme pil (SR41, LR41)
Pil ömrü Yaklaşık 1500 kez kullanım veya bekleme modu dahil
olmak üzere günde 1-2 ölçümle 1 yıl.
Boyut 12.2x2.0x1.1cm (U x G x Y)
Ağırlık: Pil dahil yaklaşık 9.5 gram
Alarm Tepe sıcaklığa ulaşıldığında yaklaşık 10 saniye sesli sinyal
Kullanım ortamı Sıcaklık: 5°C~40°C (41°F~104°F)
Nem: <95% bağıl nem.
Saklama ve taşıma koşulu Sıcaklık: -25.0°C ~ 55.0°C (-13.0°F ~ 131.0°F)
Nem: <95% bağıl nem.
Aksesuarlar pil, taşıma çantası, kullanım kılavuzu
Güvenlik sınıandırması B Tipi Uygulamalı Parça
Guidance and manufacturer’s declaration-electromagnetic emissions
The TG120 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG120 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment-guidance
RF emissions CISPR 11 Group 1 The TG120 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B The TG120 is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage uctuations/icker
emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The TG120 is intended for use in the electromagnetic environment specied below.
The customer or the user of the TG120 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance
level
Electromagnetic environment-guidance
Electro-
static discharge(ESD)
IEC61000-4-2
+ 6 kV contact
+ 8 kV air
+ 6 kV contact
+ 8 kV air
Floors should be wood, concrete or ceramic
tile. If oors are covered with synthetic mate-
rial, the relative humidity should be at least
30%
Electrical fast transient/
burst IEC61000-4-4
+ 2kV for power
supply lines
+ 1kV for input/
output lines
Not applicable
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 + 1kV line(s) to
line(s)
+ 2kV line(s) to
earth
Not applicable
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage Dips, short in-
terruptions and voltage
variations on power
supply input lines IEC
61000-4-11
<5% UT(>95%
dip in UT) for 0,5
cycle
40% UT(60% dip
in UT) for 5 cycles
70% UT(30% dip
in UT) for 25 cycles
<5% UT(>95%
dip in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of the TG120 requires continued operation
during power mains interruptions, it is recom-
mended that the TG120 be powered from an
uninterruptible power supply or a battery.
Power frequency
(50/60 Hz) magnetic
eld IEC 61000-4-8
3 A/m 3 A/m The TG120 power frequency magnetic elds
should be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey.
a. Should be less than the compliance level in each
frequency range.
b. Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorp-
tion and reection from structures, objects and people.
a. Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios,amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an
electromagnetic site survey should be considered. If the measured eld strength in the location in which
the TG120 is used exceeds the applicable RF compliance level above, the TG120 should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the TG120.
b. Over the frequency range 150 kHz to 80 MHz, eld strengths should be les than 3 V/m.
www.qutie-rossmax.com
Rossmax International Ltd.
12F., No. 189, Kang Chien Rd., Taipei, 114,
Taiwan.
Rossmax Swiss GmbH,
Tramstrasse 16, CH-9442 Berneck, Switzerland
Bu aygıt, satın alındığı tarihten itibaren 5 yıl garanti kapsamındadır; piller ve aksesuarlar dahil
değildir. Garanti, sadece bayi tarafından doldurulan ve satın alma ya da makbuz tarihini teyit eden
garanti belgesinin ibrazı koşuluyla geçerlidir. Aygıtın açılması ya da üzerinde değişiklik yapılması,
garantiyi geçersiz kılar. Hasar, kazalar veya kullanım kılavuzuna uyulmaması garanti kapsamına
girmez. Lütfen rossmax servisine başvurun.
Ürün Bilgileri
Satın alma tarihi:
Satın alındığı mağaza:
Ödenen Fiyat (Vergi hariç):
Satın alma amacı:
Bu ürün hakkındaki sohbete katıl
Burada Rossmax TG120 Termometre hakkında ne düşündüğünüzü paylaşabilirsiniz. Bir sorunuz varsa öncelikle kılavuzu dikkatlice okuyun. Kılavuz talebinde bulunmak için iletişim formumuzu kullanabilirsiniz.